5700A/5720A Series II
Operators Manual
7-22
3. Enter the ambient temperature; then press E. As the calibration check
proceeds, the display indicates the current process of the calibration check. When the
check is complete, the largest shift that is detected appears on the display.
PREV
MENU
4. You can now print a list of the proposed shifts, or quit without creating a list.
To print the report, connect a printer and set up the serial interface as described in
Chapter 6. Press the softkey under “Print Output Changes.” To return to normal
operation without printing, press the softkey under “DONE with check.”
5. Press P to exit the calibration menus.