5700A/5720A Series II
Operators Manual
1-24
5700A Series II AC Voltage Specifications: 95 % Confidence Level
Absolute Uncertainty
± 5 °C from calibration temperature
Relative Uncertainty
± 1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Range Resolution
Frequency
(Hz)
± (ppm output + µV)
2.2 mV 1 nV 10 – 20
20 – 40
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
300 k – 500 k
500 k – 1 M
400 + 4.5
170 + 4.5
85 + 4.5
300 + 4.5
700 + 7
900 + 13
1300 + 25
2800 + 25
500 + 4.5
190 + 4.5
95 + 4.5
330 + 4.5
750 + 7
1000 + 13
1500 + 25
3100 + 25
530 + 4.5
200 + 4.5
100 + 4.5
350 + 4.5
800 + 7
1050 + 13
1600 + 25
3300 + 25
550 + 4.5
210 + 4.5
105 + 4.5
370 + 4.5
850 + 7
1100 + 13
1700 + 25
3400 + 25
400 + 4.5
170 + 4.5
55 + 4.5
90 + 4.5
210 + 7
380 + 13
900 + 25
2900 + 25
500 + 4.5
190 + 4.5
60 + 4.5
100 + 4.5
230 + 7
420 + 13
1000 + 25
3200 + 25
22 mV 10 nV 10 – 20
20 – 40
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
300 k – 500 k
500 k – 1 M
400 + 5
170 + 5
85 + 5
300 + 5
700 + 7
900 + 12
1300 + 25
2800 + 25
500 + 5
190 + 5
95 + 5
330 + 5
750 + 7
1000 + 12
1500 + 25
3100 + 25
530 + 5
200 + 5
100 + 5
350 + 5
800 + 7
1050 + 12
1600 + 25
3300 + 25
550 + 5
210 + 5
105 + 5
370 + 5
850 + 7
1100 + 12
1700 + 25
3400 + 25
400 + 5
170 + 5
55 + 5
90 + 5
210 + 7
380 + 12
900 + 25
2900 + 25
500 + 5
190 + 5
60 + 5
100 + 5
230 + 7
420 + 12
1000 + 25
3200 + 25
220 mV 100 nV 10 – 20
20 – 40
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
300 k – 500 k
500 k – 1 M
400 + 13
170 + 8
85 + 8
250 + 8
700 + 25
900 + 25
1300 + 35
2800 + 80
500 + 13
190 + 8
95 + 8
280 + 8
750 + 25
1000 + 25
1500 + 35
3100 + 80
530 + 13
200 + 8
100 + 8
300 + 8
800 + 25
1050 + 25
1600 + 35
3300 + 80
550 + 13
210 + 8
105 + 8
320 + 8
850 + 25
1100 + 25
1700 + 35
3400 + 80
400 + 13
170 + 8
55 + 8
90 + 8
210 + 25
380 + 25
900 + 35
2900 + 80
500 + 13
190 + 8
60 + 8
100 + 8
230 + 25
420 + 25
1000 + 35
3200 + 80
2.2 V 1 mV 10 – 20
20 – 40
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
300 k – 500 k
500 k – 1 M
400 + 80
130 + 25
60 + 6
105 + 16
190 + 70
350 + 130
850 + 350
1700 + 850
450 + 80
140 + 25
65 + 6
110 + 16
210 + 70
390 + 130
950 + 350
1900 + 850
480 + 80
150 + 25
70 + 6
115 + 16
230 + 70
420 + 130
1000 + 350
2100 + 850
500 + 80
160 + 25
75 + 6
120 + 16
250 + 70
430 + 130
1050 + 350
2200 + 850
400 + 80
130 + 25
35 + 6
85 + 16
170 + 70
340 + 130
850 + 350
1700 + 850
450 + 80
140 + 25
40 + 6
95 + 16
190 + 70
380 + 130
950 + 350
1900 + 850
22 V 10 mV 10 – 20
20 – 40
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
300 k – 500 k
500 k – 1 M
400 + 800
130 + 250
60 + 60
105 + 160
190 + 350
400 + 1500
1050 + 4300
2300 + 8500
450 + 800
140 + 250
65 + 60
110 + 160
210 + 350
450 + 1500
1150 + 4300
2500 + 8500
480 + 800
150 + 250
70 + 60
115 + 160
230 + 350
470 + 1500
1200 + 4300
2600 + 8500
500 + 800
160 + 250
75 + 60
120 + 160
250 + 350
500 + 1500
1250 + 4300
2700 + 8500
400 + 800
130 + 250
35 + 60
85 + 160
170 + 350
400 + 1500
1000 + 4300
2200 + 8500
450 + 800
140 + 250
40 + 60
95 + 160
190 + 350
450 + 1500
1100 + 4300
2400 + 8500
± (ppm output + mV)
220 V
[2]
100 mV
10 – 20
20 – 40
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
300 k – 500 k
500 k – 1 M
400 + 8
130 + 2.5
65 + 0.8
170 + 3.5
400 + 8
1300 + 90
4300 + 90
10,500 + 190
450 + 8
140 + 2.5
70 + 0.8
190 + 3.5
450 + 8
1400 + 90
4500 + 90
11,000 + 190
480 + 8
150 + 2.5
75 + 0.8
210 + 3.5
480 + 8
1450 + 90
4600 + 90
11,300 + 190
500 + 8
160 + 2.5
80 + 0.8
220 + 3.5
500 + 8
1500 + 90
4700 + 90
11,500 + 190
400 + 8
130 + 2.5
40 + 0.8
85 + 3.5
270 + 8
1200 + 90
4200 + 90
10,500 + 190
450 + 8
140 + 2.5
45 + 0.8
95 + 3.5
300 + 8
1300 + 90
4500 + 90
11,000 + 190
1100 V
[1]
1 mV
15 – 50
50 – 1 k
340 + 16
65 + 3.5
360 + 16
70 + 3.5
380 + 16
75 + 3.5
400 + 16
80 + 3.5
340 + 16
45 + 3.5
360 + 16
50 + 3.5

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V