5700A Series II DC Current Specifications
Absolute Uncertainty
±5 °C from calibration temperature
[2]
Relative Uncertainty ±1 °C
Resolution
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Range
nA
± (ppm output + nA)
99 % Confidence Level
220 µA
2.2 mA
22 mA
0.1
1
10
45 + 10
45 + 10
45 + 100
50 + 10
50 + 10
50 + 100
55 + 10
55 + 10
55 + 100
60 + 10
60 + 10
60 + 100
24 + 2
24 + 5
24 + 50
26 + 2
26 + 5
26 + 50
µA
± (ppm output + µA)
220 mA
[1]
2.2 A
[1]
0.1
1
55 + 1
75 + 30
60 +
80 + 30
65 + 1
90 + 30
70 + 1
95 + 30
26 + 0.3
40 + 7
30 + 0.3
45 + 7
5725A Amplifier:
11 A 10 330 + 470 340 + 480 350 + 480 360 + 480 100 + 130 110 + 130
95 % Confidence Level
nA
± (ppm output + nA)
220 µA
2.2 mA
22 mA
0.1
1
10
35 + 8
35 + 8
35 + 80
40 + 8
40 + 8
40 + 80
45 + 8
45 + 8
45 + 80
50 + 8
50 + 8
50 + 80
20 + 1.6
20 + 4
20 + 40
22 + 1.6
22 + 4
22 + 40
µA
± (ppm output + µA)
220 mA
[1]
2.2 A
[1]
0.1
1
45 + 0.8
60 + 25
50 + 0.8
65 + 25
55 + 0.8
75 + 25
60 + 0.8
80 + 25
22 + 0.25
35 + 6
25 + 0.25
40 + 6
5725A Amplifier:
11 A 10 330 + 470 340 + 480 350 + 480 360 + 480 100 + 130 110 + 130
Note:
Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 mA and 2.2 mA ranges are increased
by a factor of 1.3 when supplied through 5725A terminals.
Specifications are otherwise identical for all output locations.
1. Add to uncertainty specifications:
±200 x I
2
ppm for >100 mA on 220 mA range
±10 x I
2
ppm for >1 A on 2.2 A range
2. For fields strengths >0.4 V/m but 3 V/m, add 1 % of range.

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals