5700A/5720A Series II
Operators Manual
1-38
5700A Series II AC Current Specifications: 99 % Confidence Level
Absolute Uncertainty
±5 °C from calibration temperature
[1]
Relative Uncertainty
±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Range Resolution
Frequency(
Hz)
± (ppm output + nA)
220 µA
1 nA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
650 + 30
350 + 25
120 + 20
500 + 50
1500 + 100
700 + 30
380 + 25
140 + 20
600 + 50
1600 + 100
750 + 30
410 + 25
150 + 20
650 + 50
1700 + 100
800 + 30
420 + 25
160 + 20
700 + 50
1800 + 100
450 + 30
270 + 25
110 + 20
450 + 50
1400 + 100
500 + 30
300 + 25
120 + 20
500 + 50
1500 + 100
2.2 mA 10 nA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
650 + 50
350 + 40
120 + 40
500 + 500
1500 + 1000
700 + 50
380 + 40
140 + 40
600 + 500
1600 + 1000
750 + 50
410 + 40
150 + 40
650 + 500
1700 + 1000
800 + 50
420 + 40
160 + 40
700 + 500
1800 + 1000
450 + 50
270 + 40
110 + 40
450 + 500
1400 + 1000
500 + 50
300 + 40
120 + 40
500 + 500
1500 + 1000
22 mA 100 nA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
650 + 500
350 + 400
120 + 400
500 + 5000
1500 +
10,000
700 + 500
380 + 400
140 + 400
600 + 5000
1600 + 10,000
750 + 500
410 + 400
150 + 400
650 + 5000
1700 + 10,000
800 + 500
420 + 400
160 + 400
700 + 5000
1800 + 10,000
450 + 500
270 + 400
110 + 400
450 + 5000
1400 + 10,000
500 + 500
300 + 400
120 + 400
500 + 5000
1500 +
10,000
± (ppm output + µA)
220 mA
1 µA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
650 + 5
350 + 4
120 + 4
500 + 50
1500 + 100
700 + 5
380 + 4
150 + 4
600 + 50
1600 + 100
750 + 5
410 + 4
170 + 4
650 + 50
1700 + 100
800 + 5
420 + 4
180 + 4
700 + 50
1800 + 100
450 + 5
280 + 4
110 + 4
450 + 50
1400 + 100
500 + 5
300 + 4
130 + 4
500 + 50
1500 + 100
2.2 A
10 µA
20 – 1 k
1 k – 5 k
5 k – 10 k
600 + 40
700 + 100
8000 + 200
650 + 40
750 + 100
9000 + 200
700 + 40
800 + 100
9500 + 200
750 + 40
850 + 100
10,000 + 200
600 + 40
650 + 100
7500 + 200
650 + 40
750 + 100
8500 + 200
5725A Amplifier:
11 A
100 µA
40 – 1 k
1 k – 5 k
5 k – 10 k
370 + 170
800 + 380
3000 + 750
400 + 170
850 + 380
3300 + 750
440 + 170
900 + 380
3500 + 750
460 + 170
950 + 380
3600 + 750
300 + 170
700 + 380
2800 + 750
330 + 170
800 + 380
3200 + 750
Note:
Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 µA and 2.2 mA ranges are increased by a factor of
1.3 plus 2 µA when supplied through 5725A terminals. Specifications are otherwise identical for all output locations.
1. For field strengths >0.4 V/m but 3 V/m, add 1 % of range.

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V