5700A/5720A Series II
Operators Manual
1-40
AC Current Secondary Performance Specifications and Operating Characteristics
Temperature
Coefficient
[2]
Stability ±1 °C
[1]
24 Hours
10 – 40 °C
0 – 10 °C and
40 – 50 °C
Noise and
Distortion
(Bandwidth
10 Hz – 50 kHz
<0.5V Burden)
Range
Frequency
(Hz)
± (ppm output + nA) ± (ppm output + nA)/°C
Compliance
Limits
(V rms)
Maximum
Resistive
Load
For Full
Accuracy
[3]
(Ω)
± (% output + µA)
220 µA
10 – 20
20 – 40
40 – 1 k
1 k – 5 k
5 k – 10 k
150 + 5
80 + 5
30 + 3
50 + 20
400 + 100
50 + 5
20 + 5
4 + 0.5
10 + 1
20 + 100
50 + 5
20 + 5
10 + 0.5
20 + 1
20 + 100
7 2 k
[6]
0.05 + 0.1
0.05 + 0.1
0.05 + 0.1
0.25 + 0.5
00.5 + 1
2.2 mA 10 – 20
20 – 40
40 – 1 k
1 k – 5 k
5 k – 10 k
150 + 5
80 + 5
30 + 3
50 + 20
400 + 100
50 + 5
20 + 4
4 + 1
10 + 100
50 + 400
50 + 5
20 + 4
10 + 2
20 + 100
50 + 400
7 500 0.05 + 0.1
0.05 + 0.1
0.05 + 0.1
0.25 + 0.5
00.5 + 1
22 mA 10 – 20
20 – 40
40 – 1 k
1 k – 5 k
5 k – 10 k
150 + 50
80 + 50
30 + 30
50 + 500
400 + 1000
50 + 10
20 + 10
4 + 10
10 + 500
50 + 1000
50 + 10
20 + 10
10 + 20
20 + 400
50 + 1000
7 150 0.05 + 0.1
0.05 + 0.1
0.05 + 0.1
0.25 + 0.5
00.5 + 1
Hz
± (ppm output + µA) ± (ppm output + µA) / °C
220 mA 10 – 20
20 – 40
40 – 1 k
1 k – 5 k
5 k – 10 k
150 + 0.5
80 + 0.5
30 + 0.3
50 + 3
400 + 5
50 + 0.05
20 + 0.05
4 + 0.1
10 + 2
50 + 5
50 + 0.05
20 + 0.05
10 + 0.1
20 + 2
50 + 5
7 15 0.05 + 10
0.05 + 10
0.05 + 10
0.25 + 50
00.5 + 100
2.2 A 20 – 1 k
1 k – 5 k
5 k – 10 k
50 + 5
80 + 20
800 + 50
4 + 1
10 + 5
50 + 10
10 + 1
20 + 5
50 + 10
1.4
[4]
0.5 0.5 + 100
0.3 + 500
0 1 + 1 mA