Table 3-1. Front Panel Features (continued)
M
S
Identifies a UUT full-scale endpoint for checking linearity and does not
change the output. If the output was adjusted with the rotary knob,
subsequent keyed-in output values are multiplied by a scale factor. Scaling
is deactivated by pressing
S again, or by selecting another function.
Scaling is not available for resistance outputs. See "Linearity Checking
Using Offset and Scale" in Chapter 4 for details.
N
L
Calls up a menu that allows you to specify limits beyond which the calibrator
will not operate, to protect your test equipment and personnel.
O
Z
Immediately changes the output to one tenth the reference value (not
necessarily the present output value) if the value is within performance
limits.
P
Y
Immediately changes the output to ten times the reference value (not
necessarily the present output value) if the value is within performance
limits. This key sets the calibrator to standby if this change is from below
22V to 22V or more.