Chapter 21: Specifications
DSX-5000 CableAnalyzer Module Specifications
613
Common
Mode
Rejection
40 - 20 log(f/100)
(measured to 60 dB
maximum)
37 - 20 log(f/100)
(measured to 60 dB maximum)
Tracking 0.5 dB (applicable when IL > 3 dB)
Directivity
1to 300MHz:
27 - 7log(f/100),
30 dB maximum
300 to 600 MHz:
23.7 dB
600 to 1000 MHz:
30.7 - 9 log(f/100)
(applicable when
IL > 3 dB)
1to 300MHz:
25 - 7 log(f/100), 30 dB maximum
300 to 600 MHz: 21.7 dB
600 to 1000 MHz:
27.7 - 9 log(f/100)
(applicable when IL > 3 dB)
Source Match
20 dB (applicable when IL > 3dB)
Return loss of
Termination
1to 250MHz:
20 - 15log(f/100),
25 dB maximum
250 to 600 MHz:
14 dB
600 to 1000 MHz:
21 - 9 log(f/100)
(applicable when
IL > 3 dB)
1to 250MHz:
18 - 15 log(f/100), 25 dB maximum
250 to 600 MHz: 12 dB
600 to 1000 MHz:
19 - 9 log(f/100)
(applicable when IL > 3 dB)
DSX-5000 Level V Accuracy Performance Parameters
per IEC Guidelines (continued)
Parameter Baseline Field Tester Field Tester with
Level V Permanent
Link Adapter
Field Tester with
Level V Channel
Adapter

Table of Contents