EasyManua.ls Logo

Fluke Versiv Series - Certifiber Pro OLTS Module Specifications; Certifiber Pro Power Meter Specifications

Fluke Versiv Series
700 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 21: Specifications
CertiFiber Pro OLTS Module Specifications
619
CertiFiber Pro OLTS Module Specifications
CertiFiber Pro Power Meter Specifications
Input connector
Interchangeable connector adapter
(LC standard)
Detector type
InGaAs
Wavelengths
850 nm, 1310 nm, 1550 nm
Power measurement
range
850nm: +10dBm to -65dBm
All other wavelengths: +10 dBm to -70 dBm
Display resolution
dB, dBm: 0.01
Linear (nW, μW, mW): 0.001 to 9.999, 10.00 to 99.99,
100.0 to 999.9 depending upon value
Display update rate
1 reading per second
Power measurement
uncertainty (accuracy)
 5 % 32 pW
1,2
Measurement linearity
< 0.1 dB
3
Re-calibration period
1 year
1.  5 % 100 pW at 850 nm
2. Under the following conditions:
Power level 100 μW (-10 dBm), continuous wave (CW) for absolute power at
850 nm and 1310 nm.
Divergent beam, NA = 0.20 for 50/125 μm and NA = 0.14 for 9/125 μm.
Ambient temperature 23° ± 1 °C.
SC/UPC connector with ceramic ferrule.
After a 5 minute warm-up.
Traceable to NIST.
3. -3 dBm to -55 dBm at 850 nm and 1310 nm. Ambient temperature 23 °C ± 1 °C.
After a 5 minute warm-up.

Table of Contents