Chapter 2: Get Started
Tutorial: Do a Bi-directional SmartLoop Test
53
Step 1: See How Much Memory is Available
On the home screen, tap the TOOLS icon, then tap Memory
Status.
The MEMORY STATUS screen shows these values:
The percentage of memory available
The number of test records that are saved
The number of ID sets that have been downloaded to the
tester from LinkWare PC or LinkWare Live.
The memory space taken by other files, such as the databases
for projects and limits.
Step 2: Set Up a Bi-directional SmartLoop Test
Note
You can select settings for any module that the
tester can use. It is not necessary to install a
module before you set up a test for that module.
If a module is not installed, the MODULE screen
shows to let you select the module you will use.
2-1 Attach an OptiFiber Pro module to the tester.
2-2 On the home screen, tap the test setup panel.
2-3 On the CHANGE TEST screen, tap NEW TEST.
2-4 On the TEST TYPE screen, tap SmartLoop OTDR (Auto).
2-5 On the TEST SETUP screen, tap each setting to make these
selections:
Module: This is the module you will use for the test.
(continued)