GE AKTApure - Page 120

GE AKTApure
522 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Wire stripping tool
Instruction
Follow the instructions to connect one or two external cables to the supplied D-sub connectors.
ActionStep
Open the connector housing by removing housing screw and unlatch the housing top
shell using a flat-blade screwdriver.
1
Locate the connection block mounted on the PCB board. The screw terminals has numbers
corresponding to the connector contacts.
2
54321
9876
Strip the signal cable.3
Strip-off 50 mm of the shield insulator.
Strip-off 4 mm of the single conductor insulation.
Loosen the strain relief clamp and insert the cable with the shield under the strain relief
clamp. Fasten the strain relief clamp over the cable shield.
Tip:
The connection block can be rotated inside the housing in order to position the screw ter-
minals for left side or right side cable entry.
4
Insert and fasten the single conductors in the screw terminals.5
Close the housing top shell with the latch and screw the housing together.6
System properties
Follow the instruction to update the system properties.
ActionStep
Open the system properties Edit dialog box.1
In the Component types list, click Other.2
Select the I/O-box (E9) or I/O-box 2 (E9) check box in the Component selection list. Then
click OK.
3
120 ÄKTA pure User Manual 29119969 AB
3 ÄKTA pure external modules
3.5 I/O-box E9
3.5.4 Connect external equipment to the I/O-box