System settings
Default values for digital out ports, noise reduction and configuration of analog out ports can be set.
DescriptionInstruction name
Sets the value of the signal sent out by digital port number X to
either 0 or 1. The default value is 1.
Digital out X
Filters the noise in the analog signal in port number X.Noise reduction analog in X
Enables or disables the alarm for the analog signal in port number
X. When enabled, it sets the alarm limits for the analog signal. If
the alarm is enabled and the analog signal falls outside the set
limits, an alarm will be triggered and the method will be paused.
Alarm analog in X
Enables or disables the alarm for the signal in digital port number
X. The alarm can be triggered by either of the signal values, 0 or
1. If the alarm is enabled and the condition set in ‘Value’ occurs,
an alarm will be triggered and the method will be paused.
Alarm digital in X
Enables the user to send one of the pre-defined signals (UV signal,
conductivity, temperature, pH or concentration of eluent B) to the
analog out port number X, and also to set the range of that signal.
Configure analog out X
The delay volume has to be updated if an external component is added to the flow path.
Note:
ÄKTA pure User Manual 29119969 AB 121
3 ÄKTA pure external modules
3.5 I/O-box E9
3.5.4 Connect external equipment to the I/O-box