ActionStep
Connect a 25 to 30 ml syringe to the left purge valve of the Sample pump. Make sure
that the syringe fits tightly into the purge connector.
6
Open the purge valve by turning it counter-clockwise about three-quarters of a turn.
Draw 5-10 ml of liquid slowly into the syringe with a rate of about 1 ml/s.
7
Close the purge valve by turning it clockwise. Disconnect the syringe and discard its
contents.
8
Connect the syringe to the right purge valve on the Sample pump, and repeat step 6 to
step 8.
9
ÄKTA pure User Manual 29119969 AB 171
5 Operation
5.4 Prime inlets and purge pump heads
5.4.2 Sample pump