ActionStep
In the Process Picture:6
Click on the Pumps.
Enter a low System flow (e.g., 0.2 ml/min).
Click Set flow rate.
Result: A system flow of 0.2 ml/min starts.
When buffer leaves the tubing in a continuous mode and the top part of the column is
filled with buffer, connect the tubing to the top of the column.
7
Connect a piece of tubing to the bottom of the column.8
ÄKTA pure User Manual 29119969 AB 175
5 Operation
5.5 Connect a column