DescriptionCore module
A high precision pump, which delivers buffer in purification runs.System pump P9 B or P9H B
For further information, refer to Section 2.4.1 System pumps, on page 34.
Reads the system pressure after System pump A and System pump B.Pressure monitor R9
For further information, refer to Section 2.4.12 Pressure monitors, on page68.
Mixes the buffers delivered from the system pumps to a homogeneous buffer
composition.
Mixer M9
Three Mixer chambers are available for ÄKTA pure 25, their volumes are: 0.6
ml, 1.4 ml (mounted at delivery) and 5 ml.
Three Mixer chambers are available for ÄKTA pure 150. Their volumes are:
1.4 ml (mounted at delivery), 5 ml (included in delivery), and 15 ml.
CAUTION
Risk of explosion. Do not use Mixer chamber 15 ml with
an ÄKTA pure 25 system configuration. The maximum
pressure for Mixer chamber 15 ml is 5 MPa.
For further information, refer to Section 2.4.2 Mixer, on page 38.
Directs sample onto the column.Injection valve V9-Inj or V9H-Inj
For further information, refer to Section 2.4.6 Injection valve, on page 50.
L
o
o
p
E
L
o
o
p
F
S
y
r
W
1
P
a
S
2
W
V9-Inj
ÄKTA pure User Manual 29119969 AB 19
2 The ÄKTA pure instrument
2.1 Overview illustrations