ActionStep
In the Method Editor module,5
•
click the Save the method icon
or
•
select File:Save As
Result: The Save As dialog opens.
In the Save As dialog:6
•
Select a target folder to enable the Save button.
•
Type a Name for the method.
•
Select a System from the list.
•
Click .
Result: The created method is saved in the selected folder.
Perform a System CIP
Follow the instructions below to run a System CIP method.
ActionStep
In the Method Editor module, create a System CIP method according to the instructions
above.
1
Connect bypass tubing to all selected column positions and loop positions if a loop valve
is used.
2
Prepare cleaning solutions and immerse the selected inlet tubing in the solutions.
Note:
Note that each phase uses one solution. All inlets selected in one phase should be immersed
in the same cleaning solution.
3
In the System Control module, select the created method and start the run.4
For complete cleaning of the flow path, clean the manual injection port of the Injection
valve and the pH valve manually, see the instructions below.
5
ÄKTA pure User Manual 29119969 AB 281
7 Maintenance
7.6 Maintenance when required
7.6.2 Perform System CIP