ActionStep
In the Method Editor module,7
click the Save the method icon
or
select File:Save As
Result: The Save As dialog opens.
In the Save As dialog:8
Select a target folder to enable the Save button.
Type a Name for the method.
Select a System from the list.
Click the Save button.
Result: The created method is saved in the selected folder.
Perform a Column CIP
Follow the instructions below to run a Column CIP method.
ActionStep
In the Method Editor module, create a Column CIP method according to the instruction
above.
1
Prepare cleaning solutions and immerse the selected inlets in the solutions.2
Connect the column to the selected column position.3
In the System Control module, select the created method and start the run.4
ÄKTA pure User Manual 29119969 AB 287
7 Maintenance
7.6 Maintenance when required
7.6.3 Perform Column CIP