ActionStep
In the Phase Properties tab of the Column CIP phase:4
Click Add Step to add a step.
Select the step and click Remove Step to remove a step.
To enter a value, select the cell and enter or select a new value.
In the Phase Properties tab of the Column CIP phase, click the Get Suggested Steps
button to get a suggested procedure for the selected column type. Note that this function
is not available for all column types.
5
Result: Suggested cleaning steps for the selected column type are displayed.
If several cleaning solutions are used, change settings for Inlet A and/or Inlet B . Select
one inlet for each solution. If Inlet B is used, remember to edit the values in the %B column.
6
286 ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.6 Maintenance when required
7.6.3 Perform Column CIP