Possible cause and actionProblem
Wrong wavelength for current buffer
Change wavelength or buffer.
Dirt in the UV flow cell
Clean the UV cell. See Section 7.5.1 Clean the UV flow cell, on
page 273.
Unclean optical fiber connectors
Clean the connectors. See Section 7.8.5 Replace the UV monitor
U9-M flow cell, on page 325.
Broken UV flow cell
Replace UV flow cell. See Section 7.8.5 Replace the UV monitor
U9-M flow cell, on page 325.
No light transmission
through the UV cell
Wrong wavelength for current buffer
Change wavelength or buffer.
Unclean optical fiber connectors
Clean the connectors. See Section 7.8.5 Replace the UV monitor
U9-M flow cell, on page 325.
Autozero out of accepted
range
The air intake on the rear or on the left side of the instrument
is covered
Make sure that none of the air intakes on the instrument are
covered.
Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The internal temperature of
the UV monitor is too high
No UV flow cell is attached
Attach UV cell. See Section 7.8.5 Replace the UV monitor U9-M
flow cell, on page 325.
UV flow cell is not correctly installed.
Verify that the UV cell is correctly installed. See Section 7.8.5 Re-
place the UV monitor U9-M flow cell, on page 325
The UV flow cell is broken
Replace the cell. See Section 7.8.5 Replace the UV monitor U9-M
flow cell, on page 325.
UV cell path length unread-
able
ÄKTA pure User Manual 29119969 AB 361
8 Troubleshooting
8.3 Troubleshooting: Monitors