I/O-box E9
Possible cause and actionProblem
Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35 C.
Hardware error
The internal temperature of
the I/O-box is too high
Wrong Node ID
Make sure that the Node ID is (00) for the primary box, and (01)
for a potential secondary box.
The I/O-box is not detected
by the system
Cables incorrectly connected
Make sure that the cables are connected correctly, earth to earth
etc.
The system does not detect
Digital signals correctly
Digital Out connections switched
Check that the digital Out cables are correctly connected, earth
to earth etc.
Digital In always ”0”, regard-
less of input signals
Cables incorrectly connected
Make sure that the cables are connected correctly, earth to earth
etc.
To isolate the problem:
External digital equipment
does not respond to set
changes in Digital Out
1
Connect Digital Out (e.g. pin 6) to Digital In (pin 1,2,3, or 4).
2
Change the Digital Out 1 signal between ”1” and ”0”. Verify that
Digital out follows.
3
If not: contact Service.
If it does: the problem probably lies within the connected equip-
ment.
Auto-zero on the wrong level
Reset Auto-zero.
Digital Out connection error
Make sure that Digital Out is connected correctly, earth to earth
etc.
Analog In does not measure
expected voltage
Too long cable between the external equipment and E9
Use as short cable as possible. Use shielded cable. Connect the
cable shield to the D-sub connector’s shield.
Noisy analog signal
Analog In is not calibrated
Contact Service.
Analog In does not measure
expected voltage
ÄKTA pure User Manual 29119969 AB 399
8 Troubleshooting
8.7 Troubleshooting: Other components