Possible cause and actionProblem
Analog Out is not calibrated
Contact Service.
Analog Out does not gener-
ate expected voltage
Instrument control panel
Possible cause and actionProblem
The air intake on the rear or on the right side of the instrument
is covered.
Make sure that none of the air intakes of the instrument are
covered.
Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35 C.
Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The internal temperature of
the instrument control pan-
el is too high
400 ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.7 Troubleshooting: Other components