Possible cause and actionProblem
•
Analog Out is not calibrated
Contact Service.
Analog Out does not gener-
ate expected voltage
Instrument control panel
Possible cause and actionProblem
•
The air intake on the rear or on the right side of the instrument
is covered.
Make sure that none of the air intakes of the instrument are
covered.
•
Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35 C.
•
Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The internal temperature of
the instrument control pan-
el is too high
400 ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.7 Troubleshooting: Other components