Salts and additives for hydrophobic interaction
chromatography (HIC)
The following chemicals are suitable for continuous use.
CAS no/EC noConcentrationChemical
12125-02-9/ 235-186-42 MAmmonium chloride
7783-20-2/ 231-984-13 MAmmonium sulfate
107-21-1/ 203-473-350%Ethylene glycol
56-81-5/ 200-289-550%Glycerol
Reducing agents and other additives
The following chemicals are suitable for continuous use.
CAS no/EC noConcentrationChemical
74-79-3/ 200-811-12 MArginine
100-51-6/ 202-859-92%Benzyl alcohol
3483-12-3 / 222-468-7100 mMDithioerythritol (DTE)
3483-12-3 / 222-468-7100 mMDithiothreitol (DTT)
60-00-4/ 200-449-4100 mMEthylenediaminetetraacetic acid
(EDTA)
37482-11-4/ 253-523-320 mMMercaptoethanol
7447-40-7/ 231-211-84 MPotassium chloride
Other substances
The following chemicals are suitable for continuous use.
CAS no/EC noConcentrationChemical
67-64-1/ 200-662-210%Acetone
7664-41-7/ 231-635-330%Ammonia
67-68-5/ 200-664-35%Dimethyl sulphoxide (DMSO)
75-08-1/ 200-837-320%Ethanol for long-term storage
7664-38-2/ 231-633-20.1 MPhosphoric acid
ÄKTA pure User Manual 29119969 AB 435