DescriptionInstruction name
Mixer by-pass wash is used to wash or fill the flow path (mixer
valve and tubing) that is used for sample application with system
pump. It is possible to choose a more extensive or a less extensive
wash option.
Note:
•
Pressing End during Mixer by-pass wash will terminate both
the wash and the run immediately.
Pressing Continue during Mixer by-pass wash will terminate
the wash and the run will continue from the point at which the
Mixer by-pass wash instruction was executed.
•
An instruction issued during a Mixer by-pass wash will not be
executed until the wash is completely finished and the valves
have turned back to the previous positions.
•
Mixer by-pass wash cannot be executed when the system is
in state HOLD.
Mixer by-pass wash
Loop wash is used to wash the Loop valve. It is possible to wash
a single sample application loop, all loops or only the by-pass
position. Loop wash instruction is not available, if a Column valve
is not mounted on the system.
Note:
•
Pressing End during Loop wash will terminate both the wash
and the run immediately.
Pressing Continue during Loop wash will terminate the wash
and the run will continue from the point at which the Loop wash
instruction was executed.
•
An instruction issued when a Loop wash is in progress will not
be executed until the wash is completely finished and the valves
have turned back to the previous positions.
•
Loop wash cannot be executed when the system is in state
HOLD.
Loop wash
ÄKTA pure User Manual 29119969 AB 469
9 Reference information
9.9 Manual instructions
9.9.1 Manual instructions - Pumps