DescriptionRange/UnitRun Data
The set position of the Column valve.N/AColumn position
The set flow direction position of the Column
valve V9-C, Column valve V9H-C, Column
valve V9-Cs and V9H-Cs.
N/AColumn flow direction
The system pressure signal (at the system
pumps).
-1.00 20.00 MPa (ÄKTA
pure 25)
System pressure
-1.00 5.00 MPa (ÄKTA
pure 150)
The sample pressure signal (at the sample
pump).
-1.00 10.00 MPa (ÄKTA
pure 25)
Sample pressure
-1.00 5.00 MPa (ÄKTA
pure 150)
The pre-column pressure signal.-1.00 20.00 MPa (ÄKTA
pure 25)
PreC pressure
-1.00 5.00 MPa (ÄKTA
pure 150)
The delta-column pressure signal.-1.00 20.00 MPa (ÄKTA
pure 25)
DeltaC pressure
-1.00 5.00 MPa (ÄKTA
pure 150)
The post-column pressure signal.-1.00 20.00 MPa (ÄKTA
pure 25)
PostC pressure
-1.00 5.00 MPa (ÄKTA
pure 150)
The UV absorbance signal of the U9-L monitor.-6000.000 6000.000 mAuUV
The first UV/Vis absorbance signal of the U9-
M monitor.
-6000.000 6000.000 mAUUV 1
The 2nd UV/Vis absorbance signal of the U9-
M monitor.
-6000.000 6000.000 mAUUV 2
The 3rd UV/Vis absorbance signal of the U9-
M monitor.
-6000.000 6000.000 mAUUV 3
The UV absorbance signal of the second U9-
L monitor.
-6000.000 6000.000 mAUUV 2nd
The conductivity signal.0.00 999.99 mS/cmCond
The conductivity signal as a percentage of a
set range.
0.0 100.0 %% Cond
ÄKTA pure User Manual 29119969 AB 489
9 Reference information