User Manual Druck DPI620 Genii
Executing Methods 6.12
Not all HART® devices have the same methods available.
The function, purpose and execution of each method can
be different. Methods may include the following:
• Self-test
• Loop test
• Sensor trim
• D/A trim
Execute a method as follows:
1. Select the method name. A screen opens with
information about the selected method. Four buttons
are displayed on the screen:
HELP Displays a description of the method.
ABORT Exits from the procedure.
OK Accepts inputs and proceeds to the
next step.
SWITCH APP: Returns to Druck DPI 620 Genii screen
(to change channel function settings
without interrupting the method
procedure).
Note: Some methods may make the HART® device output a
certain current.
A warning will appear on the screen before the device
enters this mode.
06 June 2013 K0541 issue 1 Page 103 of 175