User Manual Druck DPI620 Genii
Procedures (CH1): True Ohms (measure) 8.9
Do the procedure as follows:
1. Repeat procedure 8.8; in step 3 and 4 select True Ohms.
2. Check the error is within limits (Ref: Table 8-1).
Table 8-12 True Ohms (measure) error limits
Resistor
(Ω)
uncertainty
(Ω)
DPI 620 error
(Ω)
100 0.002 0.0095
06 June 2013 K0541 issue 1 Page 158 of 175