User Manual Druck DPI620 Genii
7. Check the calibration is correct
Select the TC (measure) function
8. Check the equipment gives a TC temperature that is the
same as the temperature on the reference unit ±0.1°C
(0.2°F)
Procedures (CH1): AC mV/Volts (measure) 8.13
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration for the applicable AC function.
Use the local power supply frequency.
For the AC mV (measure) function:
point 1 = 200.0 mVAC,
point 2 = 2000.0 mVAC
For the AC Volts (measure) function:
point 1 = 2.000 AC,
point 2 = 20.000 VAC
06 June 2013 K0541 issue 1 Page 165 of 175