User Manual Druck DPI620 Genii
Set the Pressure Module to Zero 3.5
SETTINGS >> ZERO >> ZERO
Use this option to write a new zero pressure value to the
pressure module in use. The sensor adjustment is permitted if
it obeys the condition that follows:
• Adjustment ≤10% FS positive pressure value (for the
Sensor).
Note: To make a temporary adjustment for zero, use the Tare
function.