User Manual Druck DPI620 Genii
Procedures (CH1/CH2): DC mV/Volts (measure) 8.4
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref: Section 8.1) to do a three-
point calibration.
(-FS, Zero and +FS) for the applicable set of ranges:
mV (measure) ranges Volts (measure) ranges
200 mV 20 V
2000 mV 30 V
4. Check the calibration is correct:
Select the applicable Millivolts or Voltage (measure)
function. (Ref: Section 2.5)
Apply the input values that are applicable to the
calibration:
mV: -2000, -1000, -200, -100, 0 (short circuit)
mV: 0, 100, 200, 1000, 2000
Volts (V): -30, -21, -20, -10, -5, 0 (short circuit)
Volts (V): 0, 5, 10, 20, 21, 30.
Check that the error is within limits (Ref: Table 8-4 or
Table 8-5).
06 June 2013 K0541 issue 1 Page 146 of 175

Table of Contents

Other manuals for GE Druck DPI 620 Genii

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Druck DPI 620 Genii and is the answer not in the manual?

GE Druck DPI 620 Genii Specifications

General IconGeneral
Protection RatingIP 54
TypeMultifunction Calibrator
Measurement ParametersPressure, Electrical
Electrical Measurement AccuracyUp to 0.005% Rdg
DisplayColor LCD touchscreen
Battery Life8 hours typical
CommunicationUSB, Bluetooth, HART
Operating Temperature-10°C to 50°C
Storage Temperature Range-20 to 70 °C

Related product manuals