Procedures (CH1): Resistance measure) 8.8
Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref:
Table 8-1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration .
Range: 0-400Ω
Nominal zero ohms:
Make a 4-wire connection to the 0
Nominal positive full-scale ohms
Make a 4-wire connection to the 400 Ω resistor
Range: 400-4k
Nominal 400Ω
Make a 4-wire connection to the 400Ω resistor
Nominal positive full-scale ohms
Make a 4-wire connection to the 4k resistor
06 June 2013 K0541 issue 1 Page 156 of 175