User Manual Druck DPI620 Genii
6.12.2 Method Example - Analog Trim
The Druck DPI 620 Genii can perform an analogue trim on
the 4 to 20 mA loop without connecting to any external
reference meters.
1. Navigate to the calibration folder
Figure 6-8 Calibration Folder
2. Select Trim analog method.
3. Follow the on-screen instructions. If channel 2 is set to
Measure Current (24V), this can be used to supply the
reference meter value.
06 June 2013 K0541 issue 1 Page 105 of 175