User Manual Druck DPI620 Genii
4. Check the calibration is correct:
• Select the applicable Current (source) function(Ref:
Section 2.4 and Section 2.9).
• Apply the values that follow:
CH1/CH2: 0.2, 6, 12, 18, 24
• Check the error is within limits (Ref: Table 8-3)
Table 8-3 Current (source) error limits
uncertainty
(mA)
DPI 620
error (mA)