User Manual Druck DPI620 Genii
4. Check the calibration is correct:
Select the applicable Current (source) function(Ref:
Section 2.4 and Section 2.9).
Apply the values that follow:
CH1/CH2: 0.2, 6, 12, 18, 24
Check the error is within limits (Ref: Table 8-3)
Table 8-3 Current (source) error limits
Source mA
Calibrator
uncertainty
(mA)
Permitted
DPI 620
error (mA)
±0.2
0.00008
0.0010
±6
0.00023
0.0016
±12
0.00044
0.0022
±18
0.0065
0.0028
±24
0.0012
0.0034
06 June 2013 K0541 issue 1 Page 145 of 175