User Manual Druck DPI620 Genii
4. Check the calibration is correct:
• Select the Resistance (source) function (Ref: Section
2.11).
• Apply the resistance values that are applicable to the
calibration (Ref: Table 8-13).
• Check the error is within limits (Ref: Table 8-13).
Table 8-13 Resistance (source) error limits
(Ω)
(mA)
uncertainty
(Ω)
DPI 620
error (Ω)
06 June 2013 K0541 issue 1 Page 160 of 175