User Manual Druck DPI620 Genii
Upgrade the HART processor Application and Boot
Loader
1. Create a folder named ‘HART’ in the root of a
USB flash memory device.
2. Copy the files ‘DK416.s19’ and ‘DK417.s19’ into
the ‘HART’ folder.
3. Put the USB flash memory drive in the USB type
A connector.
4. Select:
HART APPLICATION
5. Follow the on-screen instructions.
Note: The HART bootloader can only be upgraded as
part of an HART application upgrade.
Upgrade the CH1 FPGA
1. Create a folder named ‘FPGA’ in the root of a
USB flash memory device.
2. Copy the files ‘DK413.bin’’ into the ‘FPGA’ folder.
3. Put the USB flash memory drive in the USB type
A connector.
4. Select:
CH1 FPGA
5. Follow the on-screen instructions.
Note: The CH2 cannot be remotely upgraded .