User Manual Druck DPI620 Genii
Note: To measure or simulate TC millivolts, set the TC mV
function
1. Set the applicable software options.
2. Complete the electrical connections.
3. If necessary change the thermocouple type.
SETTINGS >> TC TYPE
4. Set CJ Compensation mode.
SETTINGS >> MANUAL CJ COMPENSATION
If Manual CJ compensation is not selected, the
internal cold junction is used to calculate the
thermocouple value.
If using an external cold junction, select check box
on MANUAL CJ COMPENSATION and enter value for
cold junction temperature.
06 June 2013 K0541 issue 1 Page 56 of 175