DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7 - 7
7-2-3 Collecting a screen capture with logs (cont’d)
Trouble image without patient information
To collect a trouble image you need to enable Transfer Images (captured without patient information)
to GE under ADMIN > System Admin.
To collect the image, press the P1 key (copy to hard drive). This places an unidentified .jpeg image in
the D:\Log \DL folder. This folder will be transferred to the back office if a transfer is scheduled.
Figure 7-4 Example of a Trouble Image Collected with P1
To collect logs using Gather Logs, the image or images will be captured in the log collection Zip file and
will be deleted from the D:\Log \DL folder.
Figure 7-5 Example of a Trouble Image Collected with Gather Logs