EasyManua.ls Logo

GE PACSystems RX7i

GE PACSystems RX7i
469 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7. PID Built-In Function Block
GFK-2950C February 2018 363
7.7 PID Example
The following PID example has a sample period of 100ms, a Kp gain of 4.00 and a Ki gain of 1.500. The
set point is stored in %R0001, the control variable is output in %AQ0002, and the process variable is
returned in %AI0003. CV Upper and CV Lower Clamps must be set, in this case to 20000 and 4000,
and an optional small Dead Band of +5 and -5 is included. The 40-word reference array starts in
%R0100. Normally, user parameters are set in the reference array, but %M0006 can be set to
re-initialize the 14 words starting at %R0102 (word 3) from constants stored in logic (a useful
technique).
The block can be switched to Manual mode with %M1 so that the Manual Command, %R113, can be
adjusted. Bits %M4 or %M5 can be used to increase or decrease %R113 and the PID CV by 1 every
100ms solution. For faster manual operation, bits %M2 and %M3 can be used to add or subtract the
value in %R2 to/from %R113 every CPU sweep. The %T1 output is on when the PID is OK.
7.7.1 Reference Array Initialization using %M00006
For details on the contents of the reference array, refer to Reference Array for the PID Function.
Word
Function
Address
Value
3
Sample Period
%R102
10
4
+ Dead Band
%R103
5
5
- Dead Band
%R104
5
6
Kp
%R105
400
7
Kd
%R106
0
8
Ki
%R107
1500
9
CV Bias
%R108
0
10
CV Upper Clamp
%R109
2000
11
CV Lower Clamp
%R110
400
12
Minimum Slew Time
%R111
0
13
Config Word
%R112
0
14
Manual Command
%R113
0
15
Control Word
%R114
0
16
Internal SP
%R115
0

Table of Contents

Other manuals for GE PACSystems RX7i

Related product manuals