ActionStep
Open the flip top of the pump head completely.2
Check that the tubing holder is adjusted to the correct position for your size
of tubing. See instructions above.
3
Place the tubing between the rotor rollers and the track, pressed against
the inner wall of the pump head.
4
ReadyToProcess WAVE 25 Operating Instructions 29009597 AD 117
5 Operation
5.1 Set up the system
5.1.5 Prepare the pump