Page #146 background image
141
Level 1
GlucoLin
Lot: 56489754
Exp.: 08
/09/03
Save?
NoNo SaveSave
70 80
17
18
19
20
16
FIGURE 6-21
The previously entered values for the
linearity material at the designated level are
displayed for confirmation.
p) To confirm the displayed values and to
save them for future comparisons to the
measured value, press Save.
FIGURE 6-22 will be displayed.
Otherwise press No. FIGURE 6-17 will be
displayed.
16 Kit name
17 Lot No.
18 Expiration Date
19 Concentration range, max. value
20 Concentration range, min. value
Please Fill
and
Insert a Cuvette
Level 1
21
FIGURE 6-22
The level of linearity material to be
tested(21) is shown on the display.
q) Fill a Cuvette from the previously
designated batch with the appropriate
level of linearity material.
r) Place the Cuvette in the Cuvette holder
and gently insert it into the measuring
position.
FIGURE 6-23 will be displayed.
21 Concentration level of Linearity material