160
11.2 Setting D/A Output
*: Selection List
When multiple circuits are measured, the items for each circuit can
be set separately. The items will be followed by a circuit No. (for
example, I1_1, I1_2, P_1, P_2).
Type Item Order Magnification
Power Voltage
(U1,U2,U3,Uave)
Current
(I1,I2,I3,I4,Iave)
Power (P,Q,S)
Power factor (PF)
Frequency (f)
Integrated power
(WP+,WP-,
WQ+,WQ-)
-- 1
Harmonic level Voltage (U1,U2,U3)
Current (I1,I2,I3,I4)
Power (P)
1 to 40 1,10,100
Harmonic
content
Voltage (U1,U2,U3)
Current (I1,I2,I3,I4)
Power (P)
1 to 40 1,10,100
Harmonic
phase angles
Voltage (U1,U2,U3)
Current (I1,I2,I3,I4)
Power (P)
1 to 40 1
Total value Voltage (U1,U2,U3)
Current (I1,I2,I3,I4)
Power (P)
-- 1
THD-F or
THD-R
Voltage (U1,U2,U3)
Current (I1,I2,I3,I4)
-- 1
• The selectable items vary depending on the wiring method.
• The full-scale ranges vary depending on the setting of "Magni-
fication" of "Harmonic level."
5-A Range
• The full-scale output (DC 5 V) varies as shown below, depend-
ing on the setting of "Magnification" of "Harmonic content."
Magnification Full-Scale Output
15 A
10 0.5 A
100 0.05 A
Magnification Full-Scale Output
1100%
10 10%
100 1%