52
4.2 Connection Methods
The wiring check result is NG.
The voltage input
is NG.
Do the voltage clips grip the wires properly?
Is the voltage cable properly inserted into
the voltage input terminal of the 3169-20/
21?
The current input
is NG.
Is the clamp-on sensor securely inserted
into the current input terminals?
Is the set current range too large for the
input level?
The voltage phase
is NG.
Are the voltage cables connected to the cor-
rect terminals?
The current phase
is NG.
Does the arrow of the clamp-on sensor point
to the load side?
Is the clamp-on sensor connected to the cor-
rect terminals?
The phase differ-
ence (I-U) is NG.
Are the voltage cables and clamp-on sensor
properly connected?
Does the arrow of the clamp-on sensor point
to the load side?
Is the power factor of the line to be mea-
sured too low, such as 0.5 or less?
The voltage bal-
ance is NG.
Does the connection method of the line to be
measured differ from that set?
Do the voltage clips grip the wires properly?
Is the voltage cable properly inserted into
the voltage input terminal?