92
_____________________________________________________________________________________________
7.4 Command Reference
______________________________________________________________________________________________
Syntax
:INITiate[:IMMediate]
Function
Changes the Idle state to the Trigger Wait state.
Note
If the unit is in the Continuous Trigger state:
(:INITIATE:CONTINUOUS 1), an execution error is issued.
If the trigger source is IMMediate, a trigger takes place immediately, putting the unit in the Idle
state.
If the trigger source is EXTernal, the unit awaits an external trigger. When it accepts a trigger, it
performs a single measurement, then goes into the Idle state.
Example
Transmission :TRIG:SOUR IMM
:INIT:CONT 0
:INIT
:FETC?
Puts the trigger system in a One-shot measurement mode, triggers once, and
reads the value.
Syntax
:TRIGger:SOURce data
data
Character data
IMMediate, EXTernal
Function
Sets the trigger source to be used.
Example
Transmission :INIT:CONT 1
:TRIG:SOUR EXT
TRG
Puts the trigger system in the Continuous state and sets the trigger source to
EXTernal. Thereafter, this function triggers once on a TRG command.
Trigger
source
Operation
IMMediate
Internal trigger
EXTernal
Triggers through the TRG terminal (external
control terminal),
TRG command, or when
M.TRIG
is pressed.
Setup of Trigger Wait state
Setup of trigger source

Table of Contents

Related product manuals