EasyManua.ls Logo

Hioki 3237 - Comparator Buzzer Sound

Hioki 3237
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
_____________________________________________________________________________________________
5.2 Comparator Buzzer Sound
______________________________________________________________________________________________
NOTE
The unit is set to "
HL
" at the factory before shipment.
5.2 Comparator Buzzer Sound
Set the buzzer to sound at comparator determination.
1
.
Press
SHIFT
.
"
SHIFT
"
lights up on the display.
2
.
Press
COMP
. The comparator buzzer sound setup screen opens, and the
current setting for the comparator buzzer flashes.
3
.Press to select the comparator determination that should activate the
buzzer.
"HL" : The buzzer beeps when the determinati on is Hi or Lo.
"In" : The buzzer beeps when the determination is In.
"
OF
" : The buzzer is not activated to beep, regardless of the comparator
result.
4
.
Press
ENT
. You are returned to the measurement screen.

Table of Contents

Related product manuals