7.4 Command Reference
______________________________________________________________________________________________
7.4.3 Specific Command Messages
Syntax
:ESE data
data
NR1 numerical data numerical data
0 to 255
Function
This command sets the available patterns of the standard event status register (SESR) to the event
status enable register 0 (ESER0).
Note
When the power is turned on, the data is reinitialized to 0.
bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0
Not used Not used
-OF +OF HI IN LO EOC
Event status enable resister 0 (ESER0)
Syntax
:ESE?
Function
The contents of ESER0 are returned as a NR1 numerical data value (0 to 255).
Syntax
:ESR0?
Function
This command returns the contents of the event status register 0 (ESR0) in NR1 numerical data
format (data) (ranging from 0 to 255), then clears the register.
Event Status Re
g
ister 0
Sets the event status enable register 0.
Read the event status enable register 0.
Read the event status register 0.

Table of Contents