76
_____________________________________________________________________________________________
7.4 Command Reference
______________________________________________________________________________________________
Syntax
[:SENSe:]FUNCtion data
data
Character data
'VOLTage[:DC]' DC voltage measurement
'VOLTage:AC' AC voltage measurement
'CURRent:DC' DC current measurement
'CURRent:AC' AC current measurement
'RESistance' 2-terminal resistance measurement
'LPResistance' Low-power 2-terminal resistance measurement
'FRESistance' 4-terminal resistance measurement
'LPRFesistance' Low-power 4-terminal resistance measurement
'CLAMp:DC' DC clamp current measurement
'CLAMp:AC' AC clamp current measurement
'FREQuency' Frequency measurement
'CONTinuity' Continuity test
'DIODe' Diode test
Function
Selects a measurement function.
Note
data must be enclosed in single quotation marks.
[:SENSe:] may be omitted.
Example
Transmission FUNC 'VOLT:DC'
Sets the DC voltage measurement function.
Syntax
[:SENSe:]FUNCtion?
data
Character data
'VOLTAGE:DC','VOLTAGE:AC','CURRENT:DC','CURRENT:AC','RESISTANCE',
'LPRESISTANCE','FRESISTANCE','LPRFESISTANCE','CLAMP:DC','CLAMP:AC',
'FREQUENCY','CONTINUITY','DIODE'
Function
Returns the currently set measurement function in a character string enclosed in single quotation
marks.
Response
syntax
Headers: ON :SENSE:FUNCTION data
Headers: OFF data
Example
Headers: ON
Transmission :SENS:FUNC?
Response :SENSE:FUNCTION 'RESISTANCE'
Headers: OFF
Transmission :SENS:FUNC?
Response 'RESISTANCE'
Measurement Composition Command
Selection of measurement functions
Inquiry concerning measurement function