93
_____________________________________________________________________________________________
7.4 Command Reference
______________________________________________________________________________________________
Syntax
:TRIGger:SOURce?
data
Character data
IMMEDIATE,EXTERNAL
Function
Inquires about the trigger source.
Response
syntax
Headers: ON TRIGGER:SOURCE data
Headers: OFF data
Example
Headers: ON
Transmission :TRIG:SOUR?
Response :TRIGGER:SOURCE IMMEDIATE
Headers: OFF
Transmission :TRIG:SOUR?
Response IMMEDIATE
The trigger source is the internal source.
Syntax
:FETCh?
data
NR3 numerical data
Function
Reads the last measurement value. Does not trigger.
Response
syntax
Headers: ON FETCH data
Headers: OFF data
Example
Headers: ON
Transmission :FETC?
Response FETCH 5.327300E-01
Headers: OFF
Transmission :FETC?
Response 5.327300E-01
The last measurement value is 0.53273. (The function and unit will vary,
depending on the type of measurement.)
Inquiry concerning trigger source
Last measurement-value read-out