Connecting and Setting Method
Setting the transmission speed (Common for USB, RS‑232C)
The instrument sets the transmission speed (baud rate) of the interface.
It is necessary to set the transmission speed when either the USB communication or the RS-232C
communication is used.
1
Press (MENU). (The setting screen appears.)
2
Select [SYST] tab.
Selection
3
Select [COM SPEED] and set SPEED.
Conrm
(or)
Cancel
Selection
Setting the controller (Computer or PLC)
Be sure to set to the following.
Start-stop synchronization
Transmission speed: 9,600 bps, 19,200 bps, 38,400 bps (Adjust to the instrument’s setting.)
Stop bit: 1
Data length: 8
Parity check: Not provided
Flow control: Not provided
IMPORTANT
The fast transmission speed (baud rate) may not be used due to a large error caused by some
computers. In that case, use with lower transmission speed.
9
Communication (RS-232C, USB)