101
10
Specications
10.1 Specications of Measurement Functions
Impedance measurement
Measurement signal Constant current AC signal
Measurement method Four-terminal pair method
Measurement terminal
structure
BNC
Measurement terminal
function
SOURCE-H terminal Current generation terminal
SOURCE-L terminal Current detection terminal
SENSE-H terminal Voltage detection terminal
SENSE-L terminal Voltage detection terminal
Measurement items Resistance (Parameter indication: R)
Reactance (Parameter indication: X)
Impedance (Parameter indication: Z)
Phase angle
(Parameter indication:
θ
)
Range structure 3 m
Ω
/10 m
Ω
/100 m
Ω
Measurement speed
setting
FAST/MED/SLOW
Display range/Resolution
3 m
Ω
range 10 m
Ω
range 100 m
Ω
range
Z Display
range
0.0000 m
Ω
to 3.6000 m
Ω
0.0000 m
Ω
to 12.0000 m
Ω
0.000 m
Ω
to 120.000 m
Ω
Resolution
0.1 µ
Ω
0.1 µ
Ω
1 µ
Ω
θ
Display
range
-180.000° to 180.000° -180.000° to 180.000° -180.000° to 180.000°
Resolution
0.001° 0.001° 0.001°
R Display
range
-0.1000 m
Ω
to 3.6000 m
Ω
-0.3000 m
Ω
to 12.0000 m
Ω
-3.000 m
Ω
to 120.000 m
Ω
Resolution
0.1 µ
Ω
0.1 µ
Ω
1 µ
Ω
X Display
range
-3.6000 m
Ω
to 3.6000 m
Ω
-12.0000 m
Ω
to 12.0000 m
Ω
-120.000 m
Ω
to 120.000 m
Ω
Resolution
0.1 µ
Ω
0.1 µ
Ω
1 µ
Ω
Frequency range 0.10 Hz to 1050 Hz
Frequency setting
resolution
0.10 Hz to 0.99 Hz 0.01 Hz step
1.0 Hz to 9.9 Hz 0.1 Hz step
10 Hz to 99 Hz 1 Hz step
100 Hz to 1050 Hz 10 Hz step
Frequency accuracy ±0.01% of setting or less