56
Checking the Judgment Result
5.5 Checking the Judgment Result
The indicator appears at the left of the parameters on the measurement screen depending on the
judgment result.
Each judgment result, and the comprehensive judgment result of all the parameters are output to
the EXT.I/O.
PASS of the comprehensive judgment result is ON (FAIL is OFF) only when all the enabled
parameters judged by the comparator are IN.
When the measured value is smaller than the upper limit value and greater than the
lower limit value.
When the measured value is greater than the upper limit value that is set.
When the measured value is smaller than the lower limit value that is set.
Judgment
result
Measurement result
Judgment
result
Output of EXT. I/O
Hi IN Lo ERR PASS FAIL
Hi Set value < Measured value Hi ON OFF OFF OFF OFF ON
Lo Set value ≤ Measured value ≤
Hi Set value
IN OFF ON OFF OFF ON OFF
Measured value < Lo Set value Lo OFF OFF ON OFF OFF ON
OverRange Hi ON OFF OFF OFF OFF ON
Measurement Error
No
judgment
OFF OFF OFF ON OFF OFF
During interruption of
measurement
No
judgment
OFF OFF OFF OFF OFF OFF