124
Discarding the Instrument
11.3 Discarding the Instrument
The instrument uses the CR2032 Coin-shaped lithium battery.
Handle and dispose of the instrument in accordance with local regulations.
Lithium battery removal
WARNING
To avoid electric shock, turn off the power switch and disconnect the power cord
and measurement cables before removing the lithium battery.
Required tools
• One Philips screwdriver (No.1)
• A pair of tweezers (to remove the lithium battery)
1
Verify that the power is off, and remove
the connection cables and power cord.
2
Remove the six screws from the sides
and one screw from the rear.
(Overhead view)
Lithium battery
3
Remove the cover.
4
Insert the tweezers between the battery
and battery holder as shown in the
diagram and lift up the battery.
IMPORTANT
Take care not to short the + and -. Doing
so may cause sparks.
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which contains Perchlorate Material - special
handling may apply. See www.dtsc.ca.gov/hazardouswaste/perchlorate