45
Preventing the Overcharge due to Measurement Signal (Voltage Limit Function)
4.6 Preventing the Overcharge due to Measurement
Signal (Voltage Limit Function)
This function prevents the battery from getting overcharged due to the applied signal when
measuring impedance. If the voltage of the object to be measured is higher compared to the set
voltage, impedance will not be measured and the message [OVER V LIMIT] will be displayed.
CAUTION
Set the voltage limit value lower than the voltage value of the measuring object’s battery
which will become overcharged. The battery may be overcharged, if the measurement
is repeated at a high voltage value setting.
1
Press (MENU). (Settings screen is displayed.)
2
Select [MEAS] tab.
Selection
3
Select [ON] or [OFF].
Conrm
(or)
Cancel
Selection
4
Customization of Measurement Conditions